This product is no longer promoted on Made-in-China.com. If you find any infringement or sensitive information of it, please contact us for handling. Thank you.

100-1000UL Micropette Mechanical Pipettes

Contact Supplier

Diamond Member Since 2016

Suppliers with verified business licenses

Manufacturer/Factory, Trading Company, Group Corporation

Basic Info.

Model NO.
HT-0334
Volume Range(UL)
0.1-2.5 and Others
Transport Package
Color Box with Export Carton
Specification
CE, ISO13485
Trademark
Dragon
Origin
Beijing, China
HS Code
9018909000
Production Capacity
10000 Sets Per Week

Product Description

Mechanical Pipettes (Adjustable & Fixed)
All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.

Features for MicroPette
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
- Manufactured from innovative materials

Features for multi-channel MicroPette
- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
- Can be used with universal style pipette tips


Calibration

Please visit our website for online calibration at hiprove.en.made-in-china.com. Through the online calibration software, users can perform simple, accurate and timely calibration with free cost, and avoid the calibration error due to non-professional operation.

We will support and help you to achieve consistently excellent results.






Specifications


 
Single-channel Adjustable Volume Pipettes
Volume Range
Increment
Test Volume
Maximum permissible systematic error (Inaccuracy)
Maximum permissible random error (Imprecision)
%
μL
%
μL
0.1-2.5μL
0.05μL
2.5μL
2.50%
0.0625
2.00%
0.05
1.25μL
3.00%
0.0375
3.00%
0.0375
0.25μL
12.00%
0.03
6.00%
0.015
0.5-10μL
0.1μL
10μL
1.00%
0.1
0.80%
0.08
5μL
1.50%
0.075
1.50%
0.075
1μL
2.50%
0.025
1.50%
0.015
2-20μL
0.5μL
20μL
0.90%
0.18
0.40%
0.08
10μL
1.20%
0.12
1.00%
0.1
2μL
3.00%
0.06
2.00%
0.04
5-50μL
0.5μL
50μL
0.60%
0.3
0.30%
0.15
25μL
0.90%
0.225
0.60%
0.15
5μL
2.00%
0.1
2.00%
0.1
10-100μL
1μL
100μL
0.80%
0.8
0.15%
0.15
50μL
1.00%
0.5
0.40%
0.2
10μL
3.00%
0.3
1.50%
0.15
20-200μL
1μL
200μL
0.60%
1.2
0.15%
0.3
100μL
0.80%
0.8
0.30%
0.3
20μL
3.00%
0.6
1.00%
0.2
50-200μL
1μL
200μL
0.60%
1.2
0.15%
0.3
100μL
0.80%
0.8
0.30%
0.3
50μL
1.00%
0.5
0.40%
0.2
100-1000μL
5μL
1000μL
0.60%
6
0.20%
2
500μL
0.70%
3.5
0.25%
1.25
100μL
2.00%
2
0.70%
0.7
200-1000μL
5μL
1000μL
0.60%
6
0.20%
2
500μL
0.70%
3.5
0.25%
1.25
200μL
0.90%
1.8
0.30%
0.6
1000-5000μL
50μL
5000μL
0.50%
25
0.15%
7.5
2500μL
0.60%
15
0.30%
7.5
1000μL
0.70%
7
0.30%
3
2-10mL
0.1mL
10mL
0.60%
60
0.20%
20
5mL
1.20%
60
0.30%
15
2mL
3.00%
60
0.60%
12
8-channel Adjustable Volume Pipettes
Volume Range
Increment
Test Volume
Maximum permissible systematic error (Inaccuracy)
Maximum permissible random error (Imprecision)
%
μL
%
μL
0.5-10μL
0.1μL
10μL
1.50%
0.15
1.50%
0.15
5μL
2.50%
0.125
2.50%
0.125
1μL
4.00%
0.04
4.00%
0.04
5-50μL
0.5μL
50μL
1.00%
0.5
0.50%
0.25
25μL
1.50%
0.375
1.00%
0.25
5μL
3.00%
0.15
2.00%
0.1
50-300μL
5μL
300μL
0.70%
2.1
0.25%
0.75
150μL
1.00%
1.5
0.50%
0.75
50μL
1.50%
0.75
0.80%
0.4
12-channel Adjustable Volume Pipettes
Volume Range
Increment
Test Volume
Maximum permissible systematic error (Inaccuracy)
Maximum permissible random error (Imprecision)
%
μL
%
μL
0.5-10μL
0

Send your message to this supplier

*From:
*To:
avatar Mr. Hank Sun
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now