Warranty: | One Year |
---|---|
Customized: | Non-Customized |
Structure: | Portable |
Material: | Plastic |
Certification: | CE, ISO, RoHS |
Application: | School, Hospital, Lab |
Customization: |
---|
Suppliers with verified business licenses
Micropipette Plus Lab Adjustable Fixed Mechanical Autoclavable Pipette
Calibration Please visit our website for online calibration at hiprove.en.made-in-china.com. Through the online calibration software, users can perform simple, accurate and timely calibration with free cost, and avoid the calibration error due to non-professional operation. We will support and help you to achieve consistently excellent results. |
Single-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.1-2.5μL | 0.05μL | 2.5μL | 2.50% | 0.0625 | 2.00% | 0.05 | |||||
1.25μL | 3.00% | 0.0375 | 3.00% | 0.0375 | |||||||
0.25μL | 12.00% | 0.03 | 6.00% | 0.015 | |||||||
0.5-10μL | 0.1μL | 10μL | 1.00% | 0.1 | 0.80% | 0.08 | |||||
5μL | 1.50% | 0.075 | 1.50% | 0.075 | |||||||
1μL | 2.50% | 0.025 | 1.50% | 0.015 | |||||||
2-20μL | 0.5μL | 20μL | 0.90% | 0.18 | 0.40% | 0.08 | |||||
10μL | 1.20% | 0.12 | 1.00% | 0.1 | |||||||
2μL | 3.00% | 0.06 | 2.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 0.60% | 0.3 | 0.30% | 0.15 | |||||
25μL | 0.90% | 0.225 | 0.60% | 0.15 | |||||||
5μL | 2.00% | 0.1 | 2.00% | 0.1 | |||||||
10-100μL | 1μL | 100μL | 0.80% | 0.8 | 0.15% | 0.15 | |||||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||||
10μL | 3.00% | 0.3 | 1.50% | 0.15 | |||||||
20-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||||
20μL | 3.00% | 0.6 | 1.00% | 0.2 | |||||||
50-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||||
100-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||||
100μL | 2.00% | 2 | 0.70% | 0.7 | |||||||
200-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||||
200μL | 0.90% | 1.8 | 0.30% | 0.6 | |||||||
1000-5000μL | 50μL | 5000μL | 0.50% | 25 | 0.15% | 7.5 | |||||
2500μL | 0.60% | 15 | 0.30% | 7.5 | |||||||
1000μL | 0.70% | 7 | 0.30% | 3 | |||||||
2-10mL | 0.1mL | 10mL | 0.60% | 60 | 0.20% | 20 | |||||
5mL | 1.20% | 60 | 0.30% | 15 | |||||||
2mL | 3.00% | 60 | 0.60% | 12 | |||||||
8-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
12-channel Adjustable Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||||
Fixed Volume Pipettes | |||||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||||
% | μL | % | μL | ||||||||
5μL | - | 5μL | 1.3% | 0.065 | 1.2% | 0.06 | |||||
10μL | - | 10μL | 0.8% | 0.08 | 0.8% | 0.08 | |||||
20μL | - | 20μL | 0.6% | 0.12 | 0.5% | 0.1 | |||||
25μL | - | 25μL | 0.5% | 0.125 | 0.3% | 0.075 | |||||
50μL | - | 50μL | 0.5% | 0.25 | 0.3% | 0.15 | |||||
100μL | - | 100μL | 0.5% | 0.5 | 0.3% | 0.3 | |||||
200μL | - | 200μL | 0.4% | 0.8 | 0.2% | 0.4 | |||||
250μL | -- | 250μL | 0.4% | 1.0 | 0.2% | 0.5 | |||||
500μL | - | 500μL | 0.3% | 1.5 | 0.2% | 1.0 | |||||
1000μL | - | 1000μL | 0.3% | 3.0 | 0.2% | 2.0 | |||||
2000μL | - | 2000μL | 0.3% | 6.0 | 0.15% | 3.0 | |||||
5000μL | - | 5000μL | 0.3% | 15 | 0.15% | 7.5 |
FAQ
Q1. Can I have a sample ?
A: Yes, we welcome sample order to test and check quality. Mixed samples are acceptable.
Q2. What about the lead time?
A:Sample needs 3-5 days, mass production time depends on your order.
Q3. Do you have any MOQ limit ?
A: Low MOQ, 1pc for sample checking is available
Q4. How do you ship the goods and how long does it take to arrive?
A: We usually ship by DHL, UPS, FedEx or TNT. It usually takes 3-5 days to arrive. Airline and sea shipping also optional.
Q5. How to proceed an order?
A: Firstly let us know your requirements or application. Secondly We quote according to your requirements or our suggestions. Thirdly customer confirms the samples and places deposit for formal order. Fourthly We arrange the production.
Q6. Is it OK to print my logo on product?
A: Yes. Please inform us formally before our production and confirm the design firstly based on our sample.
Q7: Do you offer guarantee for the products?
A: Yes, we offer 1-2 years warranty to our products.
Q8: How to deal with the faulty?
Suppliers with verified business licenses